- CNAP Machine
- Use main screen to set alarm defaults
- Put arm cuff on one arm and finger cuff on alternate limb (so when you recalibrate you continue to receive a signal)
- Set higher extremes so the alarms don't go off as often
- Respiration belt
- Place on subject so that it's not too tight but that it's firm on exhale
- GSR
- Use medical tape to close off gap on electrode around finger (create a tab to pull off easy)
- Have subjects wash their hands – no oils on their skin
- ECG
- Keep electrodes away from muscle
- Wires
- Red: lower left
- White: upper right
- Black: reference/ground
- Electrodes: exfoliate skin beneath electrodes?
*It is recommended to remove the leads from the transmitters between use
Questions & To Do:
...
- Do we want two EOG?How to calibrate CNAP with DA100CEOG measurements on both eyes or just one?
- How to add rear digital channels (buttons pad)?
- We need to redo the Valsalva setup (it asks a lot of questions that I couldn’t answer)
- Are we still going to do inspiration?Do we need to exfoliate for electrodes?
- Do we need to order any NuPrep or Elpads for electrodes? cotton pads, alcohol pads?
- What about hair on men's chests for ecg?
- Get razors?
- Get razors?
- We may need to order wrist straps (BIONOMADIX STRAP 33CM - BN-STRAP-33)
To get signal for handgrip:
...